If you’ve ever run a reliability test on your products using an Environmental Stress Screening (ESS) chamber, only to see the humidity readings spike or dip outside your pre-set threshold right when you need them most, you know how frustrating that error can be. As someone who’s spent the last 12 years building ESS test chambers—helping automotive engineers screen circuit boards, medical device makers validate implantable sensors, and aerospace teams test satellite components—I’ve seen this exact issue derail test schedules, waste valuable samples, and leave teams questioning if their chamber is even accurate. The good news? Most high humidity errors in ESS chambers aren’t random flukes. They stem from predictable, fixable issues related to how the system is designed, maintained, or operated. Let’s break down the most common culprits, explain the science behind each, and walk through what you can do to get your humidity readings back on track. ESS Test Chamber

First, it’s important to remember that ESS chambers aren’t just temperature-controlled boxes—they’re precision systems that balance temperature, humidity, and air flow to replicate real-world stressors like rapid temperature changes (thermal shock) and moisture. High humidity errors, for our purposes, are defined as readings 5% RH or more above your setpoint for a sustained period (we outline this in our standard test protocols, but that’s a separate topic). The first and most frequent cause I see, especially in older chambers or systems that haven’t had their annual calibration, is a faulty humidity sensor. But not just any faulty sensor—specifically, sensor drift, contamination, or placement error.
Let’s talk sensor drift first. Humidity sensors, most commonly capacitive sensors used in modern ESS chambers, work by measuring the change in electrical capacitance of a polymer film that absorbs and releases water vapor. Over time, this film can degrade from repeated exposure to extreme temperatures (think -40°C to 100°C, common in ESS testing), corrosive contaminants, or just thousands of humidity cycles. For example, a sensor that was calibrated to read 50% RH accurately at 25°C might drift to reading 55% RH after 2 years of use, even when the actual humidity in the chamber is 50%. What makes this tricky is that drift often happens gradually—you might not notice it at first, until you run a side-by-side test with a reference hygrometer and see the gap. I had a customer last year who was rejecting 20% of their circuit board samples for failing humidity tests, only to find their chamber sensors were 8% RH off, all due to unaddressed drift over 3 years.
Then there’s sensor contamination, which is super common in industrial ESS chambers used for testing products like lithium-ion batteries or adhesive components that outgas volatile organic compounds (VOCs). Those VOCs can stick to the sensor’s polymer film, creating a barrier that prevents it from absorbing or releasing water vapor properly. I once had a medical device customer test infusion pumps, and the plasticizer from the pump’s casing coated their sensor, making it read 15% RH higher than actual for two months before they figured it out. The fix here isn’t always a new sensor—many chambers have service kits that include sensor cleaning solutions specifically formulated for capacitive humidity sensors, so you don’t have to replace the whole unit unless the film is permanently damaged.
The third sensor-related issue is placement, which is often overlooked during chamber setup. ESS chambers have air circulation systems designed to keep the environment uniform, but if the humidity sensor is placed too close to the test product, or in a stagnant air pocket, it won’t measure the actual chamber air. For example, if you stack large metal test fixtures right in front of the sensor, you’re blocking airflow, creating a microclimate where moisture accumulates around the sensor, leading to a high reading. Or if you place the sensor near a humidity injector nozzle that’s spraying water vapor too close, you’ll get a transient high reading every time the system cycles. The rule of thumb we teach our customers is to mount the sensor at least 6 inches away from test samples, fixtures, and nozzles, in the center of the chamber’s working volume where air flow is most consistent.
The next big culprit for high humidity errors in ESS chambers is improper humidification system operation. ESS chambers use two main types of humidification: steam injection and evaporative (or water pan) humidification. Each has its own failure points, especially when paired with ESS’s rapid temperature change requirements.
Let’s start with steam injection, which is the most common for high-precision ESS chambers because it adds moisture without lowering the chamber’s temperature (a big plus for rapid temperature cycling tests). If your steam generator is malfunctioning, it can produce wet steam instead of dry steam. Wet steam has liquid water droplets in it, which add extra moisture to the chamber that’s not accounted for in the sensor setpoint. Why would steam generator produce wet steam? Two main reasons: either the boiler is scaled up (from hard water deposits that lower the boiling point, so steam isn’t fully vaporized) or the steam line has condensation buildup. For example, if you don’t drain the steam line after each test run, water accumulates, and when you inject steam for the next test, that liquid is pushed into the chamber, spiking humidity. I’ve even seen customers use tap water instead of deionized water in their steam generators—hard water has minerals that scale the boiler in weeks, leading to wet steam and constant high humidity errors. That’s a quick fix: we always recommend DI water with less than 10 microsiemens conductivity for ESS chamber humidification systems.
Evaporative humidification, while less common for rapid ESS because it can temporarily lower air temperature, has its own set of humidity issues. If the water pan at the bottom of the chamber is dirty, or if the water level is too high, it can evaporate faster than the system can remove moisture, leading to high readings. Worse, if the chamber’s refrigeration system (used to control temperature) isn’t working properly, it can’t dehumidify the air once it’s too moist. Wait a second—ESS chambers use refrigeration for cooling, right? They also use a separate dehumidification cycle, where the refrigeration coil is chilled below the dew point, so moisture condenses out of the air. If that coil is dirty with dust or debris, its cooling capacity drops, so it can’t pull enough moisture out, leaving humidity high. That’s a common maintenance issue: if you don’t clean the chamber’s air filter and refrigeration coil every 3 months, especially in dusty manufacturing environments, dehumidification efficiency drops by 30% or more, leading to consistent high humidity errors.
Another factor that’s often tied to both sensors and humidification systems is airflow imbalance in the ESS chamber. ESS relies on high-volume air circulation (usually 100 to 500 cubic feet per minute, depending on chamber size) to maintain uniform temperature and humidity, especially during rapid thermal cycles. If the air filter is clogged, or the fan motor is failing, airflow slows down, creating dead spots where moisture accumulates. For example, a customer last year had a chamber that was reading 12% RH high only in the back left corner, where their test samples were stacked. When we checked the airflow, we found the filter was 90% blocked with metal shavings from their production line, so the fan was only running at 20% capacity. Once we replaced the filter, the airflow normalized, and the humidity error was gone. What makes airflow so critical is that ESS tests often have 5 to 10 degree per minute temperature changes—if air isn’t moving fast enough, the humidity can’t stay balanced, because the air can’t carry moisture to the dehumidification system or evenly distribute it across the chamber.
Wait, I also need to mention something that’s less about equipment and more about test profile setup, which I see all the time from customers who are new to ESS. ESS tests often use a cycle called thermal ramp, where temperature goes from, say, 25°C to 85°C in 10 minutes. If you set the humidity setpoint to stay constant during that ramp, most systems struggle to keep up, because heating the air increases its capacity to hold moisture (warm air can hold much more water vapor than cold air). So when you heat the air without adjusting humidity, you can get a false reading of high humidity relative to the temperature—wait, no, actually, if you don’t add more moisture as air warms, the RH (relative humidity, which is what ESS chambers measure) drops. Wait, maybe I should correct that common mistake I see: sometimes operators accidentally set the wrong humidity parameter, like absolute humidity instead of relative humidity, and that leads to errors, but in this case, the issue is when the system can’t adjust humidity fast enough during a rapid temperature ramp. ESS chambers need to have a properly sized humidification system that can adjust moisture levels at the same rate as temperature changes. If your chamber’s humidifier is only rated for 2 liters per hour, but you’re running a 10°C per minute ramp, it can’t add moisture fast enough, but wait, no—wait, when you heat air, RH drops if moisture is constant. Maybe I mixed up the high humidity scenario here. Let me clarify: when cooling, RH can spike if the system doesn’t dehumidify enough. For example, if you’re going from 85°C to 25°C in 5 minutes, the air’s capacity to hold moisture drops drastically, so if the dehumidification system isn’t fast enough, excess moisture condenses in the chamber, leading to high RH readings. That’s a key point for ESS, because thermal ramps are core to the process—if your humidification and dehumidification systems aren’t sized for the ramp rate you’re running, you’ll get transient high humidity errors every time the chamber cools down. That’s not a sensor or sensor dirty, that’s a system mismatch between your chamber’s capacity and your test profile.
Now, let’s talk about how to diagnose which of these issues you’re dealing with, because the fix depends on the root cause. The first step is to do a simple side-by-side calibration check: use a NIST-traceable reference hygrometer, the same type we provide for annual chamber calibration, and place it in the chamber at the same height and distance from airflow as your main sensor. Run a test at a stable temperature (say 25°C, no ramp) and compare readings for 30 minutes. If the reference and chamber sensor differ by more than 2% RH, that’s a sensor issue: either drift or contamination. If they match, then the problem is with your humidification/dehumidification system or airflow.
Next, check your dehumidification performance. Set the chamber to a low humidity setpoint, like 10% RH at 25°C, and run it for an hour. If the humidity doesn’t drop to within 1% of setpoint, your dehumidification system is the culprit: check the filter, clean the refrigeration coil, test the steam generator if you use that type. If you use evaporative humidification, check the water level and pan for debris.
Then, test airflow: hold a piece of lightweight paper (like a coffee filter) in the center of the chamber. If it doesn’t stay steady and flat, the airflow is too low, which points to a clogged filter or failing fan motor.
Now, as someone who builds ESS chambers, I can also tell you that many of these errors are preventable with regular maintenance, which we outline in every chamber’s user manual, but few operators follow consistently. We provide a 12-point maintenance checklist for all our chambers that includes monthly sensor inspection, quarterly filter and coil cleaning, annual sensor calibration, and annual humidification system descaling. A customer who follows that checklist will almost never have persistent high humidity errors.
If you’ve gone through these steps and still can’t figure out the cause, or if you need help calibrating your system, replacing a faulty sensor, or upgrading your humidification system to match faster ESS ramp rates, our team of field service engineers has decades of experience troubleshooting these exact issues across all types of ESS applications. We work with automotive, aerospace, medical, and consumer electronics teams to make sure their ESS chambers are accurate and reliable, so their products get the rigorous testing they need without schedule delays.

At the end of the day, high humidity errors in ESS chambers don’t have to be a mystery. They’re almost always tied to one of a handful of avoidable issues: sensor drift or contamination, faulty humidification/dehumidification, poor airflow, or a mismatch between your chamber’s capacity and your test profile. By diagnosing the root cause correctly and staying on top of regular maintenance, you can keep your humidity readings accurate and your ESS tests running smoothly. If you’re dealing with persistent humidity errors that you can’t resolve on your own, our team is here to help you get your chamber back to peak performance—we’re ready to discuss your specific needs, whether you need maintenance support, a system upgrade, or a new ESS chamber tailored to your testing requirements.
Ozone Test Chamber References
- ASHRAE Handbook: HVAC Systems and Equipment, Chapter 24 – Humidification and Dehumidification, 2020.
- Reliability Test Methods, Part 2: Environmental Stress Screening, Institute of Environmental Sciences and Technology (IEST) Standard RP-CC001.5, 2021.
- Capacitive Humidity Sensor Technology and Application Guidelines, Honeywell Sensing and Control Technical Note, 2019.
- Environmental Chamber Maintenance Best Practices, National Institute of Standards and Technology (NIST) Special Publication 1500-30, 2022.
ALP Technology (T&M) Ltd.
With abundant experience, we are one of the most reliable ess test chamber manufacturers and suppliers in China. Please feel free to buy advanced ess test chamber for sale here from our factory. All customized products are with high quality and competitive price.
Address: No.19, Mingde Road, Xi’an, China, 710061
E-mail: info@alp-environment.com
WebSite: https://www.alp-environment.com/